Text preview for : PA-624 (B - May 2010)(4200-MAG Base).pdf part of Keithley PA-624 (B - May 2010)(4200-MAG Base) Keithley SCS 4200 PA-624 (B - May 2010)(4200-MAG Base).pdf



Back to : PA-624 (B - May 2010)(420 | Home

Models 4200-MAG-BASE, VAC-BASE, and 4225-RPM
Keithley Instruments, Inc. Probe Station Mounting Base Installation Instructions
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
www.keithley.com



Introduction

WARNING These procedures must be performed by qualified service personnel. Failure
to recognize and observe normal safety precautions may result in equipment
damage, personal injury, or even death.


This document provides installation instructions for the Keithley PreAmp Base Models 4200-MAG-BASE, 4200-
VAC-BASE, and the 4225-RPM. Read through this document and make sure all components required to complete
the installation are available before proceeding.

NOTE The Model 4200-SCS comes with a toolkit containing all the tools required to properly
install the PreAmp to the base as described in this document. Also, the Model 4225-
RPM comes with a TL-32 (3/32" hex key L-wrench) to mount the base as described in
this document.

The Model 4200-MAG-BASE and 4200-VAC-BASE are intended to provide a secure mounting surface for locating
a Model 4200-PA PreAmp or a Model 4225-RPM on a probe station platen.


PreAmp Installation
Use the following procedure for installing each PreAmp onto the MAG base.


WARNING Turn off the power from all the connected instruments and disconnect power
cords. Make sure all power is removed and stored energy in external
circuitry is discharged. Failure to remove all the power from the instruments
may result in equipment damage, personal injury, or even death.


Step 1
Install the Remote PreAmp (Model 4200-PA) onto the base by securing it with the captive screw (see Figure 1).




PA-624 Rev. B / May 2010 1
Probe station mounting base installation instructions


Figure 1: PreAmp and base assembly




Step 2
Place the PreAmp and the base assembly on the probe station platen.

Step 3
For the Model 4200-VAC-BASE only, route the vacuum hose to the hose fitting on the base (see Figure 2). Apply a
vacuum to the base. If the base does not securely hold in position, reposition the base on a clean flat surface.

NOTE To remove the vacuum base, press the Control valve pushbutton (see Figure 2).

Figure 2: Vacuum connection (Model 4200-VAC-BASE only)




2 PA-624 Rev. B / May 2010
Probe station mounting base installation instructions


Step 4
Connect the manipulator's FORCE and SENSE cables to the FORCE and SENSE triax connectors on the PreAmp
(see Figure 3).

NOTE The SENSE connection is optional.

Figure 3: Cable connections




Step 5
Connect, route, and strain relieve the Model 4200-RPC-* cable as shown in Figure 3.

Step 6 (optional)
Attach the PreAmp Ground lug to the probe station ground using the supplied Spade lug (see Figure 3). See the
Model 4200-SCS Reference manual for more details if necessary on Source Measure concepts (manual number:
4200-901-01).




PA-624 Rev. B / May 2010 3
Probe station mounting base installation instructions




Remote Pulse Measure (RPM) Installation
Use the following procedure for installing each RPM onto the MAG base.


WARNING Turn off the power from all the connected instruments and disconnect power
cords. Make sure all power is removed and stored energy in external
circuitry is discharged. Failure to remove all the power from the instruments
may result in equipment damage, personal injury, or even death.


Step 1
Install the Remote Pulse Measure unit (Model 4225-RPM) onto the base by securing it with the captive screw (see
Figure 4).
Figure 4: Captive screw and base assembly




4 PA-624 Rev. B / May 2010
Probe station mounting base installation instructions


Step 2
Fasten the RPM to the base with the socket head screw located on the side of the base (see Figure 4).
Figure 5: Socket head screw and base assembly




Step 3
Place the RPM base assembly on the probe station platen.


WARNING During final placement of the RPM base assembly, if the MAG base is used,
make sure that there is a minimum of two inches (2 in.) or five centimeters (5
cm) of space between multiple RPM base assemblies.




PA-624 Rev. B / May 2010 5
Probe station mounting base installation instructions


Step 4
For the Model 4200-VAC-BASE only, route the vacuum hose to the hose fitting on the base (see Figure 2). Apply a
vacuum to the base. If the base does not securely hold in position, reposition the base on a clean flat surface.

NOTE To remove the vacuum base, press the Control valve pushbutton (see Figure 2).

Step 5
For connection to the manipulator and instruments, see the Model 4200-SCS Reference manual for more details
(manual number: 4200-901-01).




6 PA-624 Rev. B / May 2010